JUST IN CASE . . .
ADC GUIDELINES
for the Proper
Maintenance,
Handling, Cleanup,
and
Transportation
of
Mercurial Blood
Pressure Devices
PROBLEM LIKELY CAUSE RESOLUTION
Air Bubble in Insufficient amount Add mercury by following the instructions
cartridge tube of mercury in this manual
during inflation Air trapped in To remove trapped air, tap reservoir while
reservoir system is inflated
Blood Pressure Excessive deflation Be sure to deflate the inflation system at
Reading seems High rate the recommended rate of 2-3mmHg/sec
Air flow blockage Change top cap filter
Clean or replace mercury
Blood Pressure Mercury level low Add mercury
Reading seems Low
Dirty cartridge tube Mercury has oxidized Clean tube in accordance with enclosed
leaving a silver gray instructions
residue
Mercury bouncing in Filters need Change top cap filter
tube during inflation replacements
or deflation
Mercury drops Leak at seals Make sure cartridge tube is securely seated
appear outside the Replace silicone gaskets at top and bottom
top or bottom of the of tube
cartridge tube
Mercury rises or Air flow blockage Change top cap filter
falls slowly Clean tube in accordance with enclosed
instructions
Mercury separates in Excessive deflation Be sure to deflate the inflation system at
tube during deflation rate the recommended rate of 2-3mmHg/sec
Trapped air To remove trapped air, tap reservoir while
system is inflated
IB p/n: 9353B-00 - rev 4
To Register Your Product, visit us
at www.adctoday.com
and follow the links
Consult this manual for basic service
and maintenance procedures mentioned
in this troubleshooting guide.
If any of the problems persist, contact
ADC’s Customer Service Department
Toll-Free at 1-800-ADC-2670.
Vista de pagina 0
1 2 3 4

Indice de contenidos

Pagina 1 - JUST IN CASE . .

JUST IN CASE . . .ADC GUIDELINESfor the ProperMaintenance,Handling, Cleanup, andTransportationofMercurial BloodPressure DevicesPROBLEM LIKELY CAUSE RE

Pagina 2 - 1-800-ADC-2670

INTRODUCTIONThank you for choosing an ADC DIAGNOSTIX™Mercury Sphygmomanometer. We’reproud of the care and quality that goes into the manufacture of ea

Pagina 3 - Attn: Service Department

mercury and damaged ADC DIAGNOSTIX™instrument. Please call our customer servicedepartment toll free at 1-800-ADC-2670. They will send you out a comple

Pagina 4

IF A MERCURY SPILL OCCURS . . .FAQ (FREQUENTLY ASKED QUESTIONS)Why use mercury?Mercury (Hg) is a liquid metal with very stable performance characteris

Comentarios a estos manuales

Sin comentarios